User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: ESI SR1:Stanard Resistor DATE: 13-Jun-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 3 NUMBER OF LINES: 36 CONFIGURATION: Datron 1281 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.004 IEEE [@1281]*RST 1.005 DISP Connect the UUT to the Datron in a four wire 1.005 DISP configuration. 1.006 HEAD {} 1.007 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.008 HEAD {º Resistance Calibration º} 1.009 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.010 HEAD ÍÍ Resistance Calibration ÍÍ 1.011 JMP 2.001 1.012 EVAL 2.001 MEMI Enter the nominal value for the standard resistor. 2.002 ACC Z 60.00P% 2.003 IEEE [@1281]OHMS FWR,AUTO 2.004 IEEE [@1281][D2000]RDG?[I] 2.005 MEME 2.006 MEMC Z 0.01% 3.001 HEAD {} 3.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 3.002 DISP {º THIS COMPLETES THE VERIFICATION º} 3.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 3.003 END